Where Section 17 sits and who it applies to
Annex I of Regulation (EU) 2017/745 is split into three chapters, covering the general requirements, the requirements on design and manufacture, and the information supplied with the device. Section 17 belongs to the second of these, under a heading that takes in devices incorporating electronic programmable systems as well as software that is a device in its own right.
Because of that heading, the section reaches both of the ways software comes to market. A standalone diagnostic app is caught by it, and so is the firmware running inside an infusion pump, even though almost everything else about how those two products are certified differs (our article on SaMD and SiMD covers that difference).
How you show conformity is left to you, since the MDR prescribes no form for a GSPR checklist, but what the checklist has to contain is fixed by Annex II Section 4. For each requirement the technical documentation must say whether it applies, and why not where it does not, the method used to demonstrate conformity, the harmonised standards, common specifications or other solutions applied, and the precise identity of the controlled documents holding the evidence, with a cross-reference to where they sit in the file. Measured against that, a row that names a standard and nothing else covers one of the four points at best.
17.1 Repeatability, reliability and performance
The first clause reads, “Devices that incorporate electronic programmable systems, including software, or software that are devices in themselves, shall be designed to ensure repeatability, reliability and performance in line with their intended use. In the event of a single fault condition, appropriate means shall be adopted to eliminate or reduce as far as possible consequent risks or impairment of performance.”
It asks for two different things, the first being that the software performs as claimed every time and on every configuration you say it supports. Annex II Section 6.1(b) spells out what that means for the evidence, asking that software verification and validation address all of the different hardware configurations and, where applicable, operating systems identified in the information supplied by the manufacturer. If the instructions for use list three browsers and two operating systems, the test record has to cover all of them, not just the setup the development team happens to work on.
The second duty, the single fault condition, means naming the individual faults your software can realistically meet, such as a dropped network connection, a crashed service, a corrupted or out of range input or a third party component that fails, and showing what the design does about each of them. That analysis is risk management under Sections 3 and 4 of Annex I. It belongs in the risk management file, with the GSPR row pointing to it, rather than in a separate document written to satisfy the checklist.
Machine learning adds a complication to repeatability that is worth dealing with explicitly in the file. If the same input does not always produce the same output, whether because of sampling, non-deterministic hardware or a model that changes, state the bounds you claim and how they were tested, and tie every performance figure to the model version and the test data it was measured on. Without that a reviewer has no way of telling whether the figure would be reproduced.
17.2 State of the art development and manufacture
The text of 17.2 is a single long sentence, “For devices that incorporate software or for software that are devices in themselves, the software shall be developed and manufactured in accordance with the state of the art taking into account the principles of development life cycle, risk management, including information security, verification and validation.”
Packed into that sentence are four things that must each follow the state of the art, namely the development life cycle, risk management, information security, and verification and validation. Information security appears as part of risk management here, inside 17.2, rather than being left to 17.4 or reserved for connected products, so software that never touches a network still needs its security risks assessed, even if the conclusion is that they are low.
Then there is the word manufactured, which for software means the build and release process, covering version control, the build pipeline, configuration management, release approval and the controlled distribution of each version. 17.2 expects that process to follow the state of the art as much as the design work that comes before it, and a team with a careful development process that still releases from one engineer's laptop has a gap to close.
Since the MDR never defines the state of the art, the usual way to evidence it is a harmonised standard, which under Article 8(1) gives a presumption of conformity once its reference is published in the Official Journal. That route is closed for the software life cycle, because EN 62304 is not on the list of standards harmonised under the MDR, as our IEC 62304 article explains. The references most files rely on instead are IEC 62304 for the life cycle, IEC 82304-1 for health software products, IEC 81001-5-1 for security activities across the life cycle and MDCG 2019-16 Rev.1 for cybersecurity, a guidance document that itself lists IEC 62304 and IEC 82304-1 among the relevant standards.
Most of the evidence for 17.2 is what a disciplined process produces anyway, in the form of the software development plan, the requirements and architecture, the list of third party components with their versions, the risk management file including its security risks, the verification and validation reports and the release records. The GSPR row should list each of them by document identifier rather than pointing at a folder.
17.3 Software on mobile platforms
Section 17.3 is the only clause aimed at a particular kind of platform, and it reads in full, “Software referred to in this Section that is intended to be used in combination with mobile computing platforms shall be designed and manufactured taking into account the specific features of the mobile platform (e.g. size and contrast ratio of the screen) and the external factors related to their use (varying environment as regards level of light or noise).”
For software that runs on phones or tablets, this means designing for the physical limits of the platform and for the conditions it will be used in, whether that is a small screen in bright daylight or a noisy ward. The evidence usually comes from usability engineering, carried out to IEC 62366-1 in most files, with the use environment described realistically and testing done on representative devices, together with a list of supported devices and operating system versions that the instructions for use repeat. Where the users are lay persons, Section 22.1 adds its own requirement that the device performs appropriately given their skills and the means available to them, and that the information supplied is easy for them to understand and apply.
Marking 17.3 not applicable is tempting for a web application, but it rarely survives the question of whether the application will in fact be opened on a phone. A browser-based product that works perfectly well on a small screen, with nothing in its intended purpose or instructions saying it should not be used there, is hard to defend as outside the clause. Either design and test for mobile use, or state the restriction clearly and, where you can, have the software enforce it.
17.4 Minimum requirements for hardware, networks and IT security
The last clause is also the shortest, “Manufacturers shall set out minimum requirements concerning hardware, IT networks characteristics and IT security measures, including protection against unauthorised access, necessary to run the software as intended.”
17.4 needs to be read together with its counterpart in Chapter III, Section 23.4(ab), which requires the instructions for use of a software device to state the same minimum requirements. Whatever you specify to meet 17.4 therefore has to reach the user, and a file that sets the requirements out in a design specification without carrying them into the instructions has done only half the job.
MDCG 2019-16 Rev.1 reads 17.4 as making the manufacturer responsible for setting minimum requirements for the operating environment, for those security measures that cannot be built into the device itself. The guidance defines the operating environment as any IT or network asset that interacts with the device but is not supplied by its manufacturer. It expects each requirement to rest on the device's risk assessment and the device to be as autonomous as possible in its own security, rather than depending entirely on the customer's IT, and it suggests keeping security information in an electronic form that can be updated, since threats change faster than a printed leaflet can be reissued.
For a product delivered through a browser or from the cloud, the minimum requirements will typically cover supported browsers and versions, the network conditions the software needs, what the customer's own IT has to provide, such as authentication settings or network segregation, and what the hosting environment guarantees. The same evidence usually answers two neighbouring requirements as well. Section 18.8 calls for protection, as far as possible, against unauthorised access that could stop the device working as intended, and it applies to software because Article 2(4) deems software an active device, while Section 14.2(d) requires the risks from negative interaction between software and its IT environment to be removed or reduced as far as possible.
The requirements that travel with Section 17
Section 17 is rarely answered in isolation, and MDCG 2019-16 Rev.1 maps several other Annex I requirements to software and cybersecurity. The ones below are worth placing next to the Section 17 rows in your checklist, so that the evidence cited for each stays consistent.
| GSPR | What it asks | Why it matters for software |
|---|---|---|
| 3 and 4 | A risk management system, and risk control measures chosen in a set order of priority | Single faults under 17.1 and security risks under 17.2 are outputs of this work |
| 14.2(d) | Removal or reduction of the risks from negative interaction between software and its IT environment | Covers the operating environment that 17.4 describes |
| 14.5 | Reliable and safe interoperability and compatibility with other devices or products | Interfaces, data exchange and integrations with hospital systems |
| 18.8 | Protection, as far as possible, against unauthorised access that could hamper the device | The access control and authentication behind 17.4 |
| 22.1 | Appropriate performance for lay persons, with information they can understand and apply | Patient facing apps, often on the mobile platforms of 17.3 |
| 23.4(ab) | Minimum hardware, network and IT security requirements in the instructions for use | Carries the 17.4 specification through to the user |
Writing the Section 17 rows of your GSPR checklist
Under Annex II Section 4, each row has to say whether the clause applies (with reasons if it does not), the method used, the standards, common specifications or other solutions applied, each with its edition, and the controlled documents that hold the evidence, identified by number and by location in the file. The example below shows what that looks like for 17.2 in a standalone software device.
| Column | Example entry for 17.2 |
|---|---|
| Applicability | Applies. The device is software in its own right |
| Method | Development under the documented software life cycle, with risk management that includes security risks, and verification and validation before each release |
| Standards and other solutions | IEC 62304:2006 with Amendment 1:2015, IEC 81001-5-1 and MDCG 2019-16 Rev.1, each cited by edition |
| Evidence | Software development plan, third party component list, risk management file, verification and validation report and release record, each by document identifier and location in the file |
Gaps worth checking before the file goes in
When you review your own checklist before submission, these are the points in Section 17 worth checking line by line against the evidence.
- 17.3 marked not applicable for software that can be used on a phone
- 17.4 answered in design documents but missing from the instructions for use
- Information security treated as a 17.4 topic only and absent from the 17.2 risk work
- Single fault conditions asserted in the row but never analysed in the risk file
- Standards cited without an edition, or with an edition the evidence was not produced against
- Evidence pointing to a folder or a tool rather than to a controlled document
What may change
In its proposal to amend the MDR, COM(2025) 1023 of 16 December 2025, the Commission says in the explanatory memorandum that cybersecurity will be mentioned explicitly in the Annex I requirements, and it proposes a new Article 87a obliging manufacturers to report actively exploited vulnerabilities and severe security incidents through EUDAMED. None of this is law yet, the text can change before it is adopted, and nothing in a technical file needs to change because of it today. Our article on the Cyber Resilience Act looks at the reporting proposal in more detail.
Practical steps
If you are building or reviewing the Section 17 part of a technical file, this is the order we would work in.
- Give each of 17.1 to 17.4 its own row and its own applicability decision
- List the single faults your software can meet and analyse each one in the risk file
- Assess security risks under 17.2 even if the software is never connected
- Write the 17.4 minimum requirements once and carry them into the instructions for use under 23.4(ab)
- Decide whether mobile use is intended, then design and test for it or restrict it
- Cite every standard with its edition and every piece of evidence by document identifier