What the Cyber Resilience Act is and when it applies
Regulation (EU) 2024/2847, the Cyber Resilience Act, sets cybersecurity requirements for products with digital elements, which Article 3(1) defines as a software or hardware product and its remote data processing solutions, including components placed on the market separately. Article 2(1) applies it to any such product made available on the market whose intended purpose or reasonably foreseeable use includes a data connection to a device or network.
Article 71(2) staggers the dates, so that the Act applies in general from 11 December 2027 while the reporting obligations in Article 14 have applied since 11 September 2026, and the rules on notifying conformity assessment bodies since 11 June 2026. Under Article 69(3) the reporting obligations also cover in-scope products placed on the market before 11 December 2027, which means they already reach products in use today.
Under Article 14, a manufacturer that becomes aware of an actively exploited vulnerability in its product has to send an early warning within 24 hours, a fuller notification within 72 hours and a final report no later than 14 days after a fix or mitigation is available, each going to the national computer security incident response team designated as coordinator and to ENISA. Severe incidents affecting the security of the product follow the same 24 and 72 hour steps, with the final report due within one month of the 72 hour notification.
The exclusion for medical devices
The exclusion is in Article 2(2), which reads, “This Regulation does not apply to products with digital elements to which the following Union legal acts apply: (a) Regulation (EU) 2017/745; (b) Regulation (EU) 2017/746; (c) Regulation (EU) 2019/2144.” Those are the MDR, the IVDR and the regulation on vehicle type approval.
The reasoning is set out in Recital 25, according to which the MDR and the IVDR already address cybersecurity risks, cover software and the whole life cycle, and have been backed by specific guidance on cybersecurity for medical devices since December 2019. That guidance is MDCG 2019-16, now at Rev.1 of July 2020.
For software that is a medical device, the security obligations therefore come from the MDR, mainly through Annex I Sections 17.2, 17.4, 18.8 and 23.4(ab), and they are assessed as part of the device's own conformity assessment rather than under the Act. Our guide to GSPR 17 goes through those requirements clause by clause.
The exclusion follows the product, not the company
What Article 2(2) asks is whether the MDR or the IVDR applies to the product, not whether the company making it is a medical device manufacturer. A company can place a device and a non-device product on the market under the same name, and the Act will treat the two separately. Its definition of manufacturer in Article 3(13) covers anyone who markets a product with digital elements under its own name or trademark, whether for payment, monetisation or free of charge, so giving an app away does not take it out of scope.
Everything therefore turns on qualification, meaning whether a given piece of software meets the definition of a medical device in Article 2(1) of the MDR in light of its intended purpose, with MDCG 2019-11 Rev.1 of June 2025 as the guidance (our article on Rule 11 covers the test). The table follows that decision through to the Cyber Resilience Act for the products a software device company is likely to have.
| What you place on the market | Does the MDR apply? | Cyber Resilience Act |
|---|---|---|
| Software that is a medical device in its own right | Yes | Excluded under Article 2(2)(a) |
| Software inside a device, such as embedded firmware | Yes, as part of the device | Excluded with the device |
| An accessory to a medical device | Yes, Article 1(1) covers accessories | Excluded |
| A companion or wellness app with no medical purpose, paid or free | No | In scope if it has a data connection |
| A health monitoring wearable with no medical purpose | No | In scope, and listed in Annex III as a Class I important product |
| A cloud service that does not support a product's functions | No | Outside the Act under Recital 12, though NIS2 may apply |
Wearables and cloud services
A health monitoring wearable with no medical purpose sits higher in the Act than many companies expect, because Annex III point 19 lists it as a Class I important product. Article 32(2) then requires either EU type examination or full quality assurance unless harmonised standards, common specifications or a European cybersecurity certification scheme are applied in full. A wellness wearable sold by a medical device company could therefore face a heavier conformity route under the Act than the same company's Class I device does under the MDR, where Article 52(7) lets most Class I manufacturers declare conformity themselves.
With cloud services the line is less clean than the table suggests. Recital 12 takes cloud services outside the Act where they do not support the functions of a product and points software as a service towards NIS2 instead, but Article 3(1) and (2) bring remote data processing inside the product where the manufacturer designed it, or had it designed, and the product cannot perform one of its functions without it. A back end that a non-device app depends on is therefore part of that app as far as the Act is concerned.
Modular software and the grey zone
MDCG 2019-11 Rev.1 accepts that a single piece of software can contain modules with a medical purpose alongside modules without one, such as invoicing or administrative patient records next to a clinical decision function. Modules with a medical purpose fall within the MDR, and those without one are not subject to its requirements.
The Act itself does not say how its exclusion applies to a product that is only partly a device. Until the Commission or the courts clarify it, the most defensible position is the one the MDR guidance already asks for, which is to define the intended purpose of each module, document the boundaries and dependencies between them, and decide which regime applies to every module that is made available separately as a product in its own right. If a non-medical module is sold or offered on its own, expect the Act to apply to it.
Where NIS2 fits in
NIS2, Directive (EU) 2022/2555, works differently from the Act because it regulates organisations rather than products. It requires the entities it covers to manage their own cybersecurity risks under Article 21 and to report significant incidents under Article 23, and Annex II lists manufacturers of medical devices among them, under manufacturing.
Whether it reaches a particular company depends mainly on size. Article 2(1) applies the Directive to entities of the listed types that are at least medium-sized under Commission Recommendation 2003/361/EC, subject to exceptions in Article 2(2) that apply regardless of size, and makers of devices on the public health emergency critical devices list fall under the health sector in Annex I. Being a directive, it reaches companies through each Member State's transposing law rather than directly.
For a micro or small software device company NIS2 will usually not apply, but a growing company should check the thresholds every year, because crossing them brings it into scope.
What the MDR reform proposal would add
The Commission's proposal to amend the MDR and the IVDR, COM(2025) 1023 of 16 December 2025, addresses a gap the exclusion leaves open. Recital 44 of the proposal points out that medical devices are exempt from the Act, and that security incidents which do not affect public health or patient safety are not reported under the MDR's vigilance rules at all, which it describes as an important cybersecurity gap.
To close it, the proposal inserts a new Article 87a into the MDR. Device manufacturers would report any actively exploited vulnerability in the device, and any severe incident affecting its security, to the coordinating national cybersecurity teams of the Member States where the device is available and to ENISA, through EUDAMED and within 30 days of becoming aware. Serious incident reports that also qualify would be shared with the same bodies, and the explanatory memorandum adds that cybersecurity would be named explicitly in the Annex I requirements.
All of this is still a proposal, which the Parliament and the Council can amend, and it imposes nothing until it has been adopted and starts to apply.
What to do now
Most of the work is drawing lines that ought to be drawn already, and setting up one vulnerability process that can serve whichever reporting regime ends up applying.
- List every product you place on the market under your own name, paid or free, and record the qualification decision for each
- For the device, meet the security requirements of MDR Annex I and follow MDCG 2019-16 Rev.1
- For each non-device product with a data connection, plan for the Act's requirements from 11 December 2027 and for its reporting duties, which already apply
- Check any health monitoring wearable against point 19 of Annex III before choosing its conformity route
- Document the boundaries between medical and non-medical modules, and which regime applies to each module made available separately
- Build one vulnerability intake and triage process that feeds MDR vigilance today and could meet the proposed Article 87a if it is adopted
- Check your headcount and turnover against the NIS2 thresholds each year